Neofuser Vario Plus
Seven selectable flow rates + bolus dose

Specification
- Nominal volume (ml) : 60, 100, 180, 275, 450 or 600
- Selectable flow rates (ml/h) : 0/0.5/1/1.5/2/2.5/3/3.5 or 0/1/2/3/4/5/6/7 or 0/2/4/6/8/10/12/14
- Lock-out time (min) : 3, 5, 6, 7, 8, 10, 15, 30 or 60
- Bolus volume (ml) : 0.5 (Orange), 1.0 (Blue), 2.0 (Green)
- With Luer-Lock patient connector or with NaX-patient connector ( the neuraxial application according to ISO
Description
- Neofuser Vario Plus combines the multi-rate function of Neofuser Vario with the bolus function in one model.
- The multi-rate module offers the possibility, also during drug administration, to choose from seven different flow rates. The continuous basal rate is either regulated or completely stopped by a removable key.
- The selected basal rate is confirmed by a clearly audible click sound.
- In order to avoid an overdose, the nominal bolus dose can only be retrieved at defined intervals, named lock-out-time.
- The lock-out-time is regulated by the flow rate of the additional restrictor at the entrance of the bolus reservoir.
- If the bolus is used before the lock-out-time has elapsed, a proportionally reduced bolus will be administered.
Applications
- You can select one of 7 continuous basal flows or the off mode. And when needed, you can infuse an additional bolus dose using the PCA module.